3m 5xl manual8/25/2023 He stated he tried to perform a "clear silent abort" as instructed on page 45 of the service manual but the sterilizer didn't respond. Reportedly, he was not wearing a monitor and there was no alarm in the room. Improper or Incorrect Procedure or Method (2017) Device Displays Incorrect Message (2591)ģm received information from a biomedical engineer that reported he was exposed to ethylene oxide (eo) when he was repairing a 3m steri-vac 5xl sterilizer that had stopped functioning and was displaying an e73 error. 3M HEALTH CARE 3M STERI-VAC 5XL STERILIZER/AERATOR 880.6860 STERILIZER, ETHYLENE-OXIDE GAS
0 Comments
Leave a Reply.AuthorWrite something about yourself. No need to be fancy, just an overview. ArchivesCategories |